YOKOGAWA wt1600 User Manual page 83

Digital power meter
Hide thumbs Also See for wt1600:
Table of Contents

Advertisement

[ : I N P u t ] : C U R R e n t ?
Function
Queries all settings related to the current
measurement.
Syntax
[:INPut]:CURRent?
Example
:INPUT:CURRENT? -> :INPUT:CURRENT:
TERMINAL:ELEMENT1 DIRECT;
ELEMENT2 DIRECT;ELEMENT3 DIRECT;
ELEMENT4 DIRECT;ELEMENT5 DIRECT;
ELEMENT6 DIRECT;:INPUT:CURRENT:
RANGE:ELEMENT1 5.00E+00;
ELEMENT2 5.00E+00;
ELEMENT3 5.00E+00;
ELEMENT4 5.00E+00;
ELEMENT5 5.00E+00;
ELEMENT6 5.00E+00;:INPUT:CURRENT:
SRATIO:ELEMENT1 10.0000;
ELEMENT2 10.0000;ELEMENT3 10.0000;
ELEMENT4 10.0000;ELEMENT5 10.0000;
ELEMENT6 10.0000
[ : I N P u t ] : C U R R e n t : A U T O [ : A L L ]
Function
Collectively turns ON/OFF the current auto
range of all elements.
Syntax
[:INPut]:CURRent:AUTO
[:ALL] {<Boolean>}
Example
:INPUT:CURRENT:AUTO:ALL ON
[:INPut]:CURRent:AUTO:ELEMent<x>
Function
Turns ON/OFF the current auto range of the
element or queries the current setting.
Syntax
[:INPut]:CURRent:AUTO:
ELEMent<x> {<Boolean>}
[:INPut]:CURRent:AUTO:ELEMent<x>?
<x> = 1 to 6
Example
:INPUT:CURRENT:AUTO:ELEMENT1 ON
:INPUT:CURRENT:AUTO:ELEMENT1? ->
:INPUT:CURRENT:AUTO:ELEMENT1 1
[:INPut]:CURRent:RANGe?
Function
Queries the current ranges of all elements.
Syntax
[:INPut]:CURRent:RANGe?
Example
:INPUT:CURRENT:RANGE? -> :INPUT:
CURRENT:RANGE:ELEMENT1 5.00E+00;
ELEMENT2 5.00E+00;
ELEMENT3 5.00E+00;
ELEMENT4 5.00E+00;
ELEMENT5 5.00E+00;ELEMENT6 5.00E+00
[:INPut]:CURRent:RANGe[:ALL]
Function
Collectively sets the current ranges of all
elements.
Syntax
[:INPut]:CURRent:RANGe
[:ALL] {<Current>|<Voltage>|AUTO}
IM 760101-11E
• When the crest factor is set to 3
• For a 5-A input element
<Current> = 10, 20, 50, 100, 200, 500
(mA), 1, 2, 5 (A) (when TERMinal = DIRect)
<Voltage> = 50, 100, 250, 500 (mV), 1, 2.5,
5, 10 (V) (when TERMinal = SENSor)
AUTO = Auto range
• For a 50-A input element
<Current> = 1, 2, 5, 10, 20, 50 (A) (when
TERMinal = DIRect)
<Voltage> = 50, 100, 250, 500 (mV), 1, 2.5,
5, 10 (V) (when TERMinal = SENSor)
AUTO = Auto range
• When the crest factor is set to 6
• For a 5-A input element
<Current> = 5, 10, 25, 50, 100, 250 (mA),
0.5, 1, 2.5 (A) (when TERMinal = DIRect)
<Voltage> = 25, 50, 125, 250 (mV), 0.5,
1.25, 2.5, 5 (V) (when TERMinal = SENSor)
AUTO = Auto range
• For a 50-A input element
<Current> = 0.5, 1, 2.5, 5, 10, 25 (A) (when
TERMinal = DIRect)
<Voltage> = 25, 50, 125, 250 (mV), 0.5,
1.25, 2.5, 5 (V) (when TERMinal = SENSor)
AUTO = Auto range
Example
:INPUT:CURRENT:RANGE:ALL 5A
Description The selectable range is determined by the input
element type (5A/50A) of element 1 and the
current measurement terminal setting
([:INPut]:CURRent:TERMinal:ELEMent1).
Therefore, only the elements that are set to the
same input element type and current measurement
terminal setting as element 1 are set.
[:INPut]:CURRent:RANGe:ELEMent<x>
Function
Sets the current range of the element or queries
the current setting.
Syntax
[:INPut]:CURRent:RANGe:
ELEMent<x> {<Current>|<Voltage>|
AUTO}
[:INPut]:CURRent:RANGe:ELEMent<x>?
<x> = 1 to 6
• When the crest factor is set to 3
For a 5-A input element
<Current> = 10, 20, 50, 100, 200, 500
(mA), 1, 2, 5 (A) (when TERMinal =
DIRect)
<Current > = 50, 100, 250, 500 (mV), 1,
2.5, 5, 10 (V) (when TERMinal = SENSor)
AUTO = Auto range
• For a 50-A input element
<Current> = 1, 2, 5, 10, 20, 50 (A)
(when TERMinal = DIRect)
<Current> = 50, 100, 250, 500 (mV), 1,
2.5, 5, 10 (V) (when TERMinal = SENSor)
AUTO = Auto range
5.11 INPut Group
5
5-49