YOKOGAWA wt1600 User Manual page 97

Digital power meter
Hide thumbs Also See for wt1600:
Table of Contents

Advertisement

SYNChroniz e
:MOTor?
Function
Queries all settings related to the motor
evaluation function.
Syntax
:MOTor?
Example
:MOTOR? -> :MOTOR:SPEED:
RANGE 20.0E+00;TYPE ANALOG;
PRANGE 10000.0000;PULSE 60;
SCALING 1.0000;UNIT "rpm";:MOTOR:
TORQUE:RANGE 20.0E+00;
SCALING 1.0000;UNIT "Nm";:MOTOR:PM:
SCALING 1.0000;UNIT "W";:MOTOR:
FILTER:LINE OFF;:MOTOR:POLE 2;
SSPEED I1;SYNCHRONIZE NONE
:MOTor:FILTer?
Function
Queries all settings related to the input filter.
Syntax
:MOTor:FILTer?
Example
:MOTOR:FILTER? -> :MOTOR:FILTER:
LINE OFF
: M O T o r : F I L T e r [ : L I N E ]
Function
Sets the line filter or queries the current setting.
Syntax
:MOTor:FILTer[:LINE] {OFF|
<Frequency>}
:MOTor:FILTer:LINE?
OFF = Line filter OFF
<Frequency> = 100 Hz (line filter ON, cutoff
frequency)
Example
:MOTOR:FILTER:LINE OFF
:MOTOR:FILTER:LINE? -> :MOTOR:
FILTER:LINE OFF
:MOTor:PM?
Function
Queries all settings related to the motor output.
Syntax
:MOTor:PM?
Example
:MOTOR:PM? -> :MOTOR:PM:
SCALING 1.0000;UNIT "W"
:MOTor:PM:SCALing
Function
Sets the scaling factor used for motor output
computation or queries the current setting.
Syntax
:MOTor:PM:SCALing {<NRf>}
:MOTor:PM:SCALing?
<NRf> = 0.0001 to 99999.9999
Example
:MOTOR:PM:SCALING 1
:MOTOR:PM:SCALING? -> :MOTOR:PM:
SCALING 1.0000
IM 760101-11E
<Space>
U
I
EXTernal
NONE
?
?
<x>
<x>
:MOTor:PM:UNIT
Function
Sets the unit to add to the motor output
computation result or queries the current
setting.
Syntax
:MOTor:PM:UNIT {<String>}
:MOTor:PM:UNIT?
<string> = 8 characters or less
Example
:MOTOR:PM:UNIT "W"
:MOTOR:PM:UNIT? -> :MOTOR:PM:
UNIT "W"
Description • Only the characters and symbols displayed
on the keyboard on the screen can be used.
• This command does not affect the
computation result.
:MOTor:POLE
Function
Sets the motor's number of poles or queries the
current setting.
Syntax
:MOTor:POLE {<NRf>}
:MOTor:POLE?
<NRf> = 1 to 99
Example
:MOTOR:POLE 2
:MOTOR:POLE? -> :MOTOR:POLE 2
:MOTor:SPEed?
Function
Queries all settings related to the revolution
sensor signal input.
Syntax
:MOTor:SPEed?
Example
:MOTOR:SPEED? -> :MOTOR:SPEED:
RANGE 20.0E+00;TYPE ANALOG;
PRANGE 10000.0000;PULSE 60;
SCALING 1.0000;UNIT "rpm"
:MOTor:SPEed:PRANge
Function
Sets the range of the rotating speed (pulse input
format) or queries the current setting.
Syntax
:MOTor:SPEed:PRANge {<NRf>}
:MOTor:SPEed:PRANge?
<NRf> = 0.0001 to 99999.9999
Example
:MOTOR:SPEED:PRANGE 10000
:MOTOR:SPEED:PRANGE? -> :MOTOR:
SPEED:PRANGE 10000.0000
Description This command is valid when the input type of
the revolution sensor signal (:MOTor:SPEed:
TYPE) is set to "PULSe (pulse input)."
5.14 MOTor Group
5
5-63

Hide quick links:

Advertisement

Table of Contents
loading