5.20 WAVeform Group

The commands in this group deal with the output of the retrieved waveform display data.
There are no front panel keys that correspond to the commands in this group.
:WAV eform
:
BYTeorder
:WAVeform?
Function
Queries all information about the waveform
display data.
Syntax
:WAVeform?
Example
:WAVEFORM? -> :WAVEFORM:TRACE U1;
FORMAT ASCII;START 0;END 1001
:WAVeform:BYTeorder
Function
Sets the output byte order of the waveform
display data (FLOAT format) that is transmitted
by ":WAVeform:SEND?" or queries the current
setting.
Syntax
:WAVeform:BYTeorder {LSBFirst|
MSBFirst}
:WAVeform:BYTeorder?
Example
:WAVEFORM:BYTEORDER LSBFIRST
:WAVEFORM:BYTEORDER? -> :WAVEFORM:
BYTEORDER LSBFIRST
Description This value is valid when ":WAVeform:FORMat" is
set to "{FLOat}."
IM 760101-11E
;
TRACe
<Space>
?
FORMat
<Space>
?
<Space>
?
STARt
<Space>
?
END
<Space>
?
SEND
?
LENGth
?
TRIGg er
?
SRATe
?
?
U
<x>
I
<x>
SPEed
TORQue
ASCii
FLOat
LSBFirst
MSBFirst
<NRf>
<NRf>
:WAVeform:END
Function
Sets the output end point of the waveform
display data that is transmitted by
":WAVeform:SEND?" or queries the current
setting.
Syntax
:WAVeform:END {<NRf>}
:WAVeform:END?
<NRf> to 0 to (total number of data points - 1)
Example
:WAVEFORM:END 1001
:WAVEFORM:END? -> :WAVEFORM:
END 1001
Description The ":WAVeform:LENGth?" command can be
used to query the (total number of data points).
:WAVeform:FORMat
Function
Sets the format of the waveform display data
that is transmitted by ":WAVeform:SEND?" or
queries the current setting.
Syntax
:WAVeform:FORMat {ASCii|FLOat}
:WAVeform:FORMat?
Example
:WAVEFORM:FORMAT FLOAT
:WAVEFORM:FORMAT? -> :WAVEFORM:
FORMAT FLOAT
Description For the differences in the waveform display data
output due to the format setting, see the
description for ":WAVeform:SEND?."