Make sure that the matching wall outlet receptacle is properly wired and earth-grounded.
a. Check that the line voltage agrees with the voltage listed on the name rating plate affixed
to the centrifuge.
2
Never use a three-to-two wire plug adapter.
3
Never use a two-wire extension cord or a two-wire non-grounding type of multiple-outlet
receptacle strip.
4
If there is any question about voltage, have a qualified service person measure it under load
while the drive is operating.
5
To ensure safety, the centrifuge should be wired to a remote emergency switch (preferably
outside the room where the centrifuge is housed, or adjacent to the exit from that room) in
order to disconnect the centrifuge from the main power source in case of a malfunction.

Securing the Centrifuge to the Floor

Allegra X-12 series centrifuges are certified to meet the requirements of the European CE mark. To
meet these requirements, the centrifuge must be secured to the bench using the anchoring kit
shipped with the instrument. This will prevent the centrifuge from moving in the unlikely event of
a rotor mishap.
Complete instructions for installing the anchoring kit are packaged with the hardware, which is
shipped with the centrifuge. The instructions (publication GX-TB-002) include a full-size template
to be used as a guide for drilling holes in the bench. Refer to this document for installation
instructions.

Test Run

NOTE
The centrifuge must be plugged in and the power switch turned to on position (
can be opened.
We recommend that you make a test run to ensure that the centrifuge is in proper operating
condition following shipment. See Section 3 for instructions on operating the centrifuge.
After completing the test run, return the pre-addressed warranty card included with this literature.
This will validate the centrifuge warranty and ensure your receipt of further information regarding
new accessories and/or modifications as they become available.