Troubleshooting
Retrieving Your Sample In Case of Power Failure
1
Turn the power off and disconnect the power cord from the main power source.
• No displays or LEDs are lit.
2
Make sure that the rotor is not spinning.
• No sound or vibration comes from the centrifuge.
WARNING
Never attempt to override the door interlock system while the rotor is spinning.
3
Locate the release device cap underneath the bottom center of the centrifuge (not visible from
the centrifuge front). See Figure 4-1.
Figure 4.1 Door Latch Override
1. Release Device Cap
4
Use a flat-blade screwdriver to pry the cap loose.
5
Insert the screwdriver through the two cords and pull to release.
• The latch releases and the door opens.
a. If the rotor is still spinning, close the door and wait until it stops before attempting to remove it.
WARNING
Never attempt to slow or stop the rotor by hand.
6
Reinstall the cap in the opening after sample retrieval.
4-6