Retrieving Your Sample In Case Of Power Failure - Beckman Coulter Allegra X-12 Series Instructions For Use Manual

Centrifuges
Table of Contents

Advertisement

Table 4.2 Troubleshooting Chart
Problem
Rotor cannot achieve set
speed
Door will not open
Displays are blank
Chamber does not reach
selected temperature
Chamber icing
(Allegra X-12R)

Retrieving Your Sample In Case of Power Failure

If facility power fails you will have to restart the run when the power is restored. In the event of an
extended power failure, it may be necessary to trip the door-locking mechanism manually to
remove the rotor and retrieve your sample.
GX-IM-5AB
Problem/Result
Line voltage below rating
Electrical failure
Motor failure
Rotor spinning
Power not on
Source power failure
Power not on
Electrical failure
Centrifuge cannot maintain
selected temperature for
rotor in use at speed
selected
Excess humidity in chamber
Condensation build-up
between runs
Retrieving Your Sample In Case of Power Failure
Recommended Action
Have a qualified service person measure line
voltage while the instrument is operating.
Make sure the power cord is securely
connected; call Beckman Coulter Field
Service.
Call Beckman Coulter Field Service.
Wait until the rotor stops.
Plug in the power cord; turn power on.
See
Retrieving Your Sample In Case of Power
Failure, below.
Plug in the power cord; turn power on.
Make sure the power cord is securely
connected; call Beckman Coulter Field
Service.
• Precool or prewarm rotors before running
at low or high temperatures.
• Precool rotor chamber by running a
30-minute cycle at the required
temperature with the speed set at
2000 rpm (Allegra X-12R).
• Make sure the air intake vent is clear.
• If using multiwell-plates for runs
exceeding 2 hours in a warm and/or
humid environment, speed reduction may
be required to maintain low sample
temperature.
Wipe out moisture before each run.
• Leave door open between runs.
• Set temperature to a setting higher than
ambient temperature.
• Turn centrifuge power off.
4
Troubleshooting
4-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Allegra x-12 seriesAllegra x-15r

Table of Contents