Other Possible Problems - Beckman Coulter Allegra X-12 Series Instructions For Use Manual

Centrifuges
Table of Contents

Advertisement

Troubleshooting

Other Possible Problems

Table 4.1 Diagnostic Message Chart (Continued)
Diagnostic Number/
Message
I1–
Rotor load is severely
Imbalance, see
out of balance
manual
Rotor tie-down is not
properly fastened
Centrifuge is
misaligned (tilted)
Centrifuge was moved
during operation
Drive error
(mechanical damage)
I2–
System imbalance
Imbalance error, see
error
manual
Other Possible Problems
Operating problems that may not be indicated by diagnostic messages are described in
along with probable causes, listed in the probable order of occurrence, and corrective actions.
Perform the recommended corrective action in sequence, as listed. If you are unable to correct the
problem, call Beckman Coulter Field Service.
4-4
Problem
Result
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Run shuts down
with maximum
brake
Recommended Action
• Make sure that tubes or bottles are
balanced and loaded symmetrically in the
rotor as described in the applicable rotor
manual.
• With swinging bucket rotors, remove the
buckets and clean and lubricate the pivot
pin sockets where the buckets contact as
described in the applicable rotor manual.
Unlubricated pivot pin sockets can
prevent the buckets from reaching
horizontal position, which can cause
imbalance.
• Press CE to clear message.
Make sure rotor lid is properly aligned and tie-
down T-bar is securely fastened. Press CE to
clear message.
Level the centrifuge on the bench or table.
Press CE to clear message.
After the rotor comes to a complete stop,
open and close the centrifuge door, then
restart. Press CE to clear message.
Call Beckman Coulter Field Service.
Press CE to clear message, then restart run. If
problem persists, call Beckman Coulter Field
Service.
Table
4.2,
GX-IM-5AB

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Allegra x-12 seriesAllegra x-15r

Table of Contents