Electrical Requirements - Beckman Coulter Allegra X-12 Series Instructions For Use Manual

Centrifuges
Table of Contents

Advertisement

Installation

Electrical Requirements

Ambient temperatures during operation should not be lower than 10°C (50°F) or higher than
35°C (95°C). Relative humidity should not exceed 75% (noncondensing).
The centrifuge must have adequate air ventilation to ensure conformance to local
requirements for vapors produced during operation.
Figure 2.1 Centrifuge Dimensions
7.6
3.0
Electrical Requirements
208-V, 60-Hz centrifuge
200-V, 50/60-Hz centrifuge
230-V, 50-Hz centrifuge
To reduce the risk of electrical shock, this centrifuge comes with a 2.5-m (8-ft) three-wire electrical
cord (attached to the power connector at the rear of the instrument) and plug to connect the
centrifuge to earth-ground. The plug for use with 208-VAC, 60-Hz centrifuges is the NEMA 6-15P;
a NEMA 6-15R socket should be available. In regions where the instrument is supplied with an
unterminated cord, a plug that meets local electrical and safety requirements must be supplied.
(Contact your local Beckman Coulter Service Office for specific information regarding these
requirements.)
2-2
cm
in.
34.3
13.5
76.2
30.0
187–229 VAC, 9 A, 60 Hz
180–220 VAC, 10 A, 50/60 Hz
207–253 VAC, 8 A, 50 Hz
96.5
38.0
62.2
24.5
7.6
3.0
GX-IM-5AB