Operation

Summary of Run Procedures

Summary of Run Procedures
For runs at other than room temperature, refrigerate or warm the rotor beforehand for fast equilibration.

Manual Run

1
Turn the power switch on (
2
Press
DOOR
3
Install the rotor according to the applicable rotor manual, then close the chamber door.
4
Press
ROTOR
5
Press
RPM/RCF
toggle between RPM and RCF modes.
6
Press
TIME
twice to toggle between hold (continuous) run and set time mode.
7
Press
TEMP °C
(Allegra X-12R only).
8
Press
ACCEL
(
1
SLOW
9
Press
DECEL
(
) through
0
OFF
10
Check that all parameters are correct and that the door is closed (door LED will be lit).
Press
ENTER/SAVE
11
Wait for the set time to count down to zero, or end the run by pressing
12
When the rotor stops (a tone sounds), press
3-2
).
I
to open the chamber door; lift the door open.
, then press
again or use the arrow keys to select the rotor.
ROTOR
, then use the keypad to enter the run speed. Press
, then use the keypad to enter the run time (to 99 hrs, 59 minutes); or press
, then use the keypad to enter the required run temperature –10 to +40°C
, then use the keypad or arrow keys to enter the selected acceleration rate number,
) through
(
).
10
MAX
, then use the keypad or arrow keys to enter the selected deceleration rate number,
(
).
10
MAX
, then press
START
(within 5 seconds).
to open the chamber door; lift the door.
DOOR
a second time to
RPM/RCF
TIME
STOP.
GX-IM-5AB