IEEE 488 Configuration Settings
Standard Units, Only
1. Ensure IEEE 488 Interface is installed.
2. Turn the ChartScan system's power "OFF."
3. Ensure the COMM SELECT micro-switch (on the
rear-panel DIP switch) is in the "1" (up) position.
This selects IEEE 488.
4. Adjust micro-switches 4 through 8 (for address
setting), see following table for standard units.
Note: Early production ChartScan units do not have a Calibration Enable/Disable Pushbutton. If you have an
early production unit, refer to the right-hand DIP switch illustration, above.
IEEE 488 Settings (Standard Units)
Selection
Micro-
Switch
#
COMM
1
SELECT
HANDSHAKE
2,3
(H/S)
IEEE 488
4,5
ADDRESS
Decimal Values
6,7,8
RS-232 or
9
RS-422
3-12
Setting
0 – Serial Communication; RS-232 or
RS-422 depending on switch 9.
1- Selects IEEE 488.
No effect when switch 1 is set to "1".
0, 0 - 0
0, 1- 8
1, 0- 16
1, 1- 24
0,0,0- 0
0,0,1- 1
0,1,0- 2
0,1,1- 3
1,0,0- 4
1,0,1- 5
1,1,0- 6
1,1,1- 7
No effect when switch 1 is set to "1".
Early Production Units, Only
1. Ensure IEEE 488 Interface is installed.
2. Turn the ChartScan system's power "OFF."
3. Ensure the COMM SELECT micro-switch (on the
rear-panel DIP switch) is in the "0" (down) position.
This selects IEEE 488.
4. Adjust micro-switches 4 through 8 (for parameter
preferences), see following table for early production
units.
IEEE 488 Settings (Early Production Units)
Selection
Micro-
Switch
#
COMM
1
SELECT
HANDSHAKE
2,3
(H/S)
IEEE 488
4,5
ADDRESS
Decimal Values
6,7,8
CALIBRATION
9
MODE
Setting
0- selects IEEE 488 interface card.
1- Selects RS-232/422.
No effect when switch 1 is set to "0".
0, 0 - 0
0, 1- 8
1, 0- 16
1, 1- 24
0,0,0- 0
0,0,1- 1
0,1,0- 2
0,1,1- 3
1,0,0- 4
1,0,1- 5
1,1,0- 6
1,1,1- 7
0- Calibration Mode Disabled
1- Calibration Mode Enabled (normal
position is "0" disabled)