Omega Engineering OMB-CHARTSCAN 1400 User Manual page 163

Portable data recorder
Table of Contents

Advertisement

Appendix A
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Digital Input Stamping (
scan as it occurs in real time.
The digital input state consists of 8 bits (
where "1" indicates digital input "on" and "0" indicates digital input "off."
In binary mode, 2 bytes will be appended as follows: Byte 2: 00h,
In ASCII mode, 8 bytes will be appended as follows:
Where
Note: If Alarm Stamping (
EXAMPLES
Without Digital Input Stamping (default):
PRINT#1, "OUTPUT07;I#0X"
PRINT#1, "OUTPUT07;R1X"
PRINT#1, "ENTER07"
INPUT A$
....+0234.20-0019.40+0001.40+0023.60
With Digital Input Stamping:
PRINT#1, "OUTPUT07;I#1X"
PRINT#1, "OUTPUT07;R1X"
PRINT#1, "ENTER07"
INPUT A$
....+0234.20-0019.40+0001.40+0023.60 036 000
ChartScan User's Manual
- Digital Input Stamping
I#
System
Deferred
Specifies digital input stamping as enabled,or disabled.
I#state
) command, when enabled, will append the digital input state to the end of each
I#
= User Terminator, and
D
XXX
) is also enabled, the Digital Input Stamp (
A#
Alarm Stamp.
either of the following values: the default of
stamping, or
to enable digital stamping.
1
through
). Each bit has an assigned binary value of "1" or "0,"
I8
I1
I8 - I1
DXXXD000
= the decimal equivalent of
' Disable digital input stamp
' Configure and start an acquisition
' Request the next scan in the buffer
' Retrieve it
'
' Enable digital input stamping
' Configure and start an acquisition
' Request the next scan in the buffer
' Retrieve it
state
, which disables digital input
0
I8 - I1
.
) will be appended after the
I#
API Commands
may have
A-37

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omb-chartscan-1400

Table of Contents