Omega Engineering OMB-CHARTSCAN 1400 User Manual page 154

Portable data recorder
Table of Contents

Advertisement

API Commands
TYPE
EXECUTION
SYNTAX
DESCRIPTION
Note: This command is only for advanced users who perform their own calibration. It is not necessary for
To use this command the unit must be in Calibration Mode. The End Calibration Mode (
used to take the unit out of Calibration Mode. Once this command has been executed the unit will no longer be
in Calibration Mode and the Enter Calibration Mode command (
Chapter 6, Calibration contains additional information.
EXAMPLE
PRINT#1, "OUTPUT07;K12345X"
PRINT#1, "OUTPUT07;H17X"
PRINT#1, "OUTPUT07;G17,1X"
PRINT#1, "OUTPUT07;G17,1X"
PRINT#1, "OUTPUT07;J1,2,90.0X"
PRINT#1, "OUTPUT07;EX"
A-28
- End Calibration Mode
E
Calibration
Immediate
Terminate Calibration Mode
E
normal, everyday operation.
) must be issued to re-enter Calibration Mode.
K
' Enter Calibration Mode
' Apply precision voltage
' Will calculate offsets for card #2
' Wait for command to complete
' Apply precision voltage of -50.0 mV
' Calibrate negative gain for TC for card #2
'Wait for command to complete
' Apply precision voltage of +50.0 mV
' Calibrate positive gain for TC for card #2
' Wait for command to complete
' Apply temperature of 90.0 °C
' Calibrate cold junction offset for channel 1 on card 1
' Wait for command to complete
' Remove precision voltage
' End Calibration Mode
Appendix A