Appendix A
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Set Burst Mode Frequency (
High-speed, single-channel mode (burst mode). See Measuring Modes section of Appendix C for more
information.
For AC measurements where the RMS value is to be returned via the
responsibility to program a frequency which will yield a sufficient whole even number of samples per line cycle in
order to compute the RMS accurately. In general for AC measurements, the frequency should equal (2n * 60) Hz
where n is a positive integer.
EXAMPLE
PRINT#1, "OUTPUT07;F#15000.0X"
PRINT#1, "OUTPUT07;F#100.5X"
ChartScan User's Manual
- Set Burst Mode Frequency
F#
System
Deferred
Select burst mode sampling frequency where
F#freq
where
38.5
) command is used to select the channel sampling frequency when in the
F#
<
<
Hertz (default).
freq
20000.0
command, it is the user's
U16
' Select 15 KHz sampling frequency
' Select 100.5 Hz sampling frequency
API Commands