Gain Calibration Of Main Unit - Omega Engineering OMB-CHARTSCAN 1400 User Manual

Portable data recorder
Table of Contents

Advertisement

Gain Calibration of Main Unit

Required Equipment*
- Volts DC Calibrator
- 22-2 Twisted pairs (to connect calibrator to calibration card)
- Digital Voltmeter, or Digital Multimeter (used for verification of calibration voltages)
- Calibration Card (installed in bottom slot, slot #1, of main unit)
*
see required specifications on page 6-2
Offset calibration of main unit must be completed prior to calibrating the main unit for
gain.
Use approved ESD precautions, including static-free work area and grounded wrist strap,
when handling circuit boards and electronic components. Failure to do so could cause
equipment damage due to electrostatic discharge.
6-8 Calibration
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
ChartScan User's Manual