Appendix A
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Restore Factory Defaults (
when it left the factory. Issuing this command will cause the unit to load the factory default configuration into the
working configuration. The factory default configuration is defined in the Configuring Power-Up State section of
Appendix C.
This command may not be issued while an acquisition is configured since it will cause certain settings upon
which the acquisition depends to change. Issuing this command while an acquisition is configured will cause a
Conflict Error to occur.
Note: Issuing the
EXAMPLE
PRINT#1,"OUTPUT07;Q?T?Y?X"
PRINT#1,"ENTER07"
INPUT A$
PRINT#1,"OUTPUT07;*FX"
PRINT#1,"OUTPUT07;Q?T?Y?X"
PRINT#1,"ENTER07"
INPUT A$
ChartScan User's Manual
- Restore Factory Settings
*F
System
Immediate
Restore unit to factory configuration
*F
) command may be used to restore the unit to the same configuration it had
*F
command will cause any previously defined configuration to be irrevocably lost. The
*F
calibration factors, however, will remain intact.
' Retrieve some current configuration settings
' Get the settings
' Screen shows
Q1,7,7,7,0 T0,3,0,0 Y000100,
001000,000000
' Restore factory default configuration
' Retrieve some current configuration settings
' Get the settings
' Screen shows
Q1,0,0,0,0 T0,0,0,0 Y000000,
000000,000000
API Commands
A-15