API Commands
TYPE
EXECUTION
SYNTAX
DESCRIPTION
The Power-On Reset (
data in data buffers and configuration are erased.
Note: Because the
The IEEE 488 bus commands Device Clear (DCL and SDC) do not have this effect. They clear only the
command input buffer, the output queue, and any pending commands.
Note: Issuing the Power-Up Default Settings (
EXAMPLE
PRINT#1,"OUTPUT07;*RX"
SLEEP 4
WHILE (S% and 4) = 0
PRINT #1, "SPOLL07"
INPUT#2,S%
WEND
A-18
*R
System
Immediate
Power-On Reset. Equivalent to a hardware reset.
*R
) command has the same effect on the unit as removing and re-applying power. All
*R
command performs a full power-on reset, 3 or 4 seconds are required before you can
*R
communicate with the unit. A total of at least 5 seconds is required before normal operations can take
place.
command will cause any previously defined configuration to be irrevocably lost. The calibration factors for
the unit and its slaves, however, will remain intact.
- Power-On Reset
) command and then power-cycling the unit, or issuing the
*S1
' Restore power-on settings to the unit
' Wait 4 seconds to restore communication
' Keep checking the serial poll response of the unit until it is ready,
meaning that bit 4 is asserted
Appendix A
*R