Table 2-1: Final Test And - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Table 2-1: Final Test and Calibration Values
TEST Values
RANGE
STABIL
PRESS
SAMP FL
PMT
UV LAMP
STR. LGT
DRK PMT
DRK LMP
SLOPE
OFFSET
HVPS
DCPS
RCELL TEMP
BOX TEMP
PMT TEMP
IZS TEMP
PMT Volts
SO
Conc
2
PMT Volts
SO
Conc
2
(table continued)
Analyzer Operator Manual, 02164, Rev. G
2
Observed
Units
Value
PPB
PPB
in-Hg
CC / MIN
mV
mV
PPB
mV
mV
mV
V
mV
o
C
o
C
o
C
o
C
Electric Test & Optic Test
Electric Test
mV
PPB
Optic Test
mV
PPB
2-12
Nominal Range
50 - 20,000
0.1 - 2
25 - 35
650 ± 10%
0 - 5000
2000 - 4000 typical
<100
-50 - +200
-50 - +200
1.0 ± 0.3
<250
550 - 900 constant
2500 ± 200
50 ± 1
8-50
7 ± 1
50 ± 0.3
2000 ± 1000
1000 ± 500
2000 ± 1000
1000 ± 500
Reference Section
5.3.4
5.2.1, 9.1.1,
Table 9-1, 9.2.5
9.1.1, 9.3.6,
Table 9-1
9.2.2, 9.3.6,
Table 9-1
9.4.1
9.4.2
Table 9-1
9.4.3, Table 9-1
Table 9-1
7.10
7.10
9.3.10
9.3.5
9.3.7
9.3.4.1
9.3.9
9.5.4
9.1.3.2
9.1.3.2
9.1.3.3
9.1.3.3