Teledyne 6400A Instruction Manual page 61

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Example 2: to perform zero calibration adjust once per day retarding 15 minutes everyday starting
at 11:30 pm, 12/20/98.
1) MODE: ZERO
2) TIMER ENABLE: ON
3) STARTING DATE: 12/20/98
4) STARTING TIME: 23:30
5) DELTA DAYS: 0
6) DELTA TIME: 23:45
7) DURATION: 15.0 MINUTES
8) CALIBRATE: ON
Example 3: to perform zero-span calibration check once per day at 10:30 pm and zero
calibration adjust once per week starting at 11:30 PM, 12/20/98.
1. Select any one of SEQx and program as example 1.
2. Select any other SEQx and program as following; Always avoid setting two or more
sequences at the same time of the day.
1) MODE: ZERO
2) TIMER ENABLE: ON
3) STARTING DATE: 12/20/98
4) STARTING TIME: 23:30
5) DELTA DAYS: 7
6) DELTA TIME: 00:00
7) DURATION: 15.0 MINUTES
8) CALIBRATE: ON
Analyzer Operator Manual, 02164, Rev. G
2
6-5