Calibration Quality; Table 7-16: Calibration - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
6400A with IZS option can be used to calibrate zero/span of one Range #, but not both. Other Range
# should be calibrated using external zero/span gas through sample port. Pressing CALZ (for zero)
or CALS (for span) keys will lead to show #1 and #2 the same way as CAL key except CALZ is
dedicated for zero air calibration while CALS is dedicated for span gas calibration. Zero/span valve
option is also treated same as IZS option except external zero/span gas are used for both Range#.

7.10 Calibration Quality

After Zero/Span is complete, it is very important to check the QUALITY of the calibration. The
calibration of the 6400A involves balancing several sections of electronics and software to
achieve an optimum balance of accuracy, noise, linearity and dynamic range.
The following procedure compares the Slope and Offset parameters in the equation used to
compute the SO
calculation see Section 5.2.2.5.
The slope and offset parameters are similar to the span and zero pots on an analog instrument.
Just as in the analog instrument, if the slope or offset get outside of a certain range, the
instrument will not perform as well.
The offset value gives information about the background signal level. Check the observed offset
value against the factory value in Table 2-1. If significantly higher check Section 9.1.6.
Increasing readings are a predictor of problems.
Table 7-16: Calibration Quality Check
Step Number
1.
2.
After the above procedure is complete, the 6400A is ready to measure sample gas.
Analyzer Operator Manual, 02164, Rev. G
2
concentration. For an explanation of the use of these terms in the concentration
2
Action
Scroll the TEST
function menu
until the SLOPE
is displayed.
Scroll the TEST
function menu
until the
OFFSET is
displayed.
Comment
Typical SLOPE value for SO
in this range, check Section 9.1.6. If the SLOPE value is in the
acceptable range the instrument will perform optimally.
Typical number is between 50mV and 250mV which is mainly
the optical system background. If the OFFSET value is outside
this range, check Section 9.1.6.
7-25
is 1.0 ± 0.3. If the value is not
2