Trace Oxygen Analyzer
Use the ◄►arrow keys to position the blinking over the Logout
function, and press Enter to Log out. The screen will display the
message:

4.3.5 System Self-Diagnostic Test

The Model 3000TB has a built-in self-diagnostic testing routine.
Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the
test the status of each function is displayed on the screen, either as OK
or as a number between 1 and 3. (See System Self Diagnostic Test in
chapter 5 for number code).
The self diagnostics are run automatically by the analyzer
whenever the instrument is turned on, but the test can also be run by the
operator at will. To initiate a self diagnostic test during operation:
Press the System button to start the System function.
Use the ◄►arrow keys to blink More, then press Enter.
Use the ◄►arrow keys again to move the blinking to the Self–
Test function. The screen will follow the running of the diagnostic.
Teledyne Analytical Instruments
Protected Until
Password Reentered
TRAK/HLD Auto—Cal
PSWD Logout More
Version Self—Test
RUNNING DIAGNOSTIC
Testing Preamp — 83
Operation
45