Warning Commands And Messages; Table 6-11: Rs-232 W - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
The T command lists TEST messages. Examples of the T command are:
T LIST
T LIST ALL
T RCELLTEMP
T SO2CONC
T LAMPRATIO

6.8.4 WARNING Commands And Messages

Table 6-11: RS-232 Warning Messages
Name
WSYSRES
WRAMINIT
WSAMPFLOW
WSAMPPRESS
WUVLAMP
WPMTTEMP
WSHUTTTER
WRCELLTEMP
WBOXTEMP
WIZSTEMP
WDYNZERO
WDYNSPAN
WHVPS
WVFDET
WDCPS
Analyzer Operator Manual, 02164, Rev. G
2
Lists test messages currently enabled with T SET
Lists all test messages
Prints the temperature of the reaction cell
Prints SO
concentration message
2
Prints Lamp Ratio
Message
SYSTEM RESET
RAM INITIALIZED
SAMPLE FLOW WARNING
SAMPLE PRESSURE WARNING
UV LAMP WARNING
PMT TEMP WARNING
SHUTTER WARNING
RCELL TEMP WARNING
BOX TEMP WARNING
IZS TEMP WARNING
CANNOT DYN ZERO
CANNOT DYN SPAN
HVPS WARNING
V/F NOT INSTALLED
DCPS WARNING
6-18
Description
Analyzer was reset/powered on
RAM was erased
Sample flow out of spec.
Sample pressure below 15" Hg or above
35" Hg
UV lamp output is below 600mV or
above 4995mV or PMT above 4995mV
PMT temperature too high or too low
Shutter not functioning
Reaction cell temp. out of spec.
Box temperature too high or too low
IZS temp. out of spec.
Dynamic zero cal. out of spec.
Dynamic span cal. out of spec.
HVPS too high or too low
A/D board not installed or broken
DC power supply output below
2300mV or above 2700mV