Figure 7-3: Diagramf P - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
1. Zero calibration procedure is identical to Table 7-11.
2. Generate the SO
2
Calculate the expected concentration as follows.
×
3
P
10
=
×
[
SO
]
382 10
.
+
2
F
F
c
4
[SO
]
= SO
gas concentration, ppm
2
2
P
= permeation flow rate at the specific temperature, ug·SO
F
= carrier flow rate over the permeation tube, SLPM (25
c
F
= diluent air flow rate, SLPM (25
d
Figure 7-3: Diagram F Permeation Calibration System
Analyzer Operator Manual, 02164, Rev. G
2
concentration to approximately 80% of the URL of the range.
×
4
--------------------------------------- Equation 7.2
7-17
o
C, 760 mmHg)
/min.
2
o
C, 760 mmHg)