Table 7-8: Activitym - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Table 7-8: Activity Matrix for Calibration Equipment & Supplies
Equipment/supplies
Recorder
Sample line and
manifold
Calibration
equipment
Working standard
SO
cylinder gas or
2
SO
permeation tube
2
Zero air
Record form
Audit equipment
Analyzer Operator Manual, 02164, Rev. G
2
Acceptance limits
Compatible with
output signal of
analyzer; min. chart
width of 150 mm (6
in) is recommended
Constructed of PTFE
or glass
Meets guidelines of
reference 1 and
Section 2.3.2 (Q.A.
Handbook)
Traceable to NIST-
SRM meets limits in
traceability protocol
for accuracy and
stability (Section
2.0.7, Q.A.
Handbook)
Clean dry ambient
air, free of
contaminants that
cause detectable
response with the
SO
analyzer.
2
Develop standard
forms
Must not be the
same as used for
calibration
7-11
Frequency and method
of measurement
Check upon receipt
Check upon receipt
See Section 2.3.9 (Q.A.
Handbook)
Analyzed against NIST-
SRM; see protocol in
Section 2.0.7, Q.A.
Handbook
See Section 2.9.2 (Q.A.
Handbook)
N/A
System must be checked
out against known
standards
Action if requirements
are not met
Return equipment to
supplier
Return equipment to
supplier
Return equipment/
supplies to supplier or
take corrective action
Obtain new working
standard and check for
traceability
Obtain air from another
source or regenerate.
Revise forms as
appropriate
Locate problem and
correct or return to
supplier