Table 7-3: Entere; Table 7-4: Manual Span - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Enter the expected SO
2
Table 7-3: Enter Expected Span Gas Concentrations Procedure
Step Number
Action
1.
Press CAL-
CONC
2.
Press ENTR
3.
Press EXIT
Table 7-4: Manual Span Calibration Procedure - Span Gas Through Sample Port
Step Number
Action
1.
Press CAL
2.
Wait 10 min
3.
Press SPAN
4.
Press ENTR
5.
Press EXIT
Analyzer Operator Manual, 02164, Rev. G
2
span gas concentration:
Comment
This key sequence causes the 6400A to prompt for the
expected SO
value by pressing the key under each digit until the expected
value is set. This menu can also be entered from CALS.
ENTR stores the expected SO
Returns instrument to SAMPLE mode.
Comment
The 6400A enters the calibrate mode. External span gas should
be fed to the sample port.
Wait for reading to stabilize at span value
If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.
Pressing ENTR actually changes the calibration equations and
causes the instrument to read the SO
The 6400A returns to sampling. Immediately after calibration,
data is not added to the DAS averages.
7-5
concentration. Enter the SO
2
span value.
2
span concentration
2
span concentrations.
2