8 MAINTENANCE

8.1 Maintenance Schedule

Table 8-1: Preventative Maintenance Schedule
Item
TEST functions
Zero/Span Calibration
Zero/Span checks
Particulate Filter
Zero air canister
Zero Air DFU filter
Reaction cell cleaning
Sample Flow
Pneumatic Lines
Factory Calibration
Leak Check
Pump Diaphragm
Analyzer Operator Manual, 02164, Rev. G
2
The operations outlined in this chapter are to be
performed by qualified maintenance personnel only.
Maintenance Interval
Check every month
Weekly or as needed
Daily
Weekly as needed
Refill every 3 months
Change every 12 months
Clean annually and clean as necessary
Check every 6 months
Examine every 12 months, clean if
necessary
Calibrate each year or after repairs
Check every year
Change annually
8-1
NOTE
Reference Section
Table 9-1
Section 7
Section 7, Table 6-2
Figure 8-1
Section 8.4
Section 8.4
Section 8.6, Figure 8-3
Figure 8-3, Section 9.2.2
Figure 8-3
Section 9.1.6
Section 8.8