Teledyne 6400A Instruction Manual page 133

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No
Signal
19
ST_DIAG_MODE
20
ST_POWER_OK
21
ST_SYSTEM_OK
22
ST_HVPS_ALARM
23
PMT_SIGNAL
24
RCELL_TEMP
25
BOX_TEMP
26
IZS_TEMP
27
PMT_TEMP
28
DCPS_VOLTAG
29
SAMPLE_PRES
30
SAMPLE_FLOW
31
HVPS_VOLTAG
32
DAC_CHAN_0
33
DAC_CHAN_1
(table continued)
Analyzer Operator Manual, 02164, Rev. G
2
Control
Description
YES
Status Bit - In Diagnostic mode
Logic High = 6400A in Diagnostic mode
Logic Low = Not in Diagnostic mode
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
YES
Status Bit - System OK
Logic High = No instrument warning present
Logic Low = 1 or more alarm present
YES
Status Bit - HVPS alarm
Logic High = HVPS out of spec
Logic Low = HVPS within spec
NO
Current PMT voltage. Same as PMT voltage in TEST
menu. Bi-polar, typically in 0-5000mV range. A
constant value of 5000mV indicates offscale.
NO
Reaction Cell temperature. 3500 mV for 50
NO
Box Temperature. Typically 1800 mV for 25
NO
IZS permeation tube oven temp. Typically 3500 mV
o
for 50
C.
NO
PMT cold block temperature. Typically 4300 mV for
o
7
C.
NO
DC power supply composite voltage output.
Typically 2500 mV.
NO
Sample pressure in mV. Typical sea level value =
4300mV for 29.9" HG-A.
NO
Sample flow in mV. Typical value is about 3200 mV
for 650cc/min.
NO
HVPS programming voltage. Output of HVPS is
1000x value present.
NO
Output of SO
NO
Output of SO
9-15
(DAC0/RANGE1) in mV.
2
(DAC1/RANGE2) in mV.
2
o
C.
o
C.