Table 6-8: Rs-232 C - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Table 6-8: RS-232 Command Summary
Terminal Mode Editing Keys
V MODE
BS
EXC
^R
^E
CR
^C
Computer Mode Editing Keys
LF
^T
Security Features
LOGON password
LOGOFF
General Output Message Format
Reporting of status messages for use as an audit trail is one of the two principal uses for the RS-
232 interface. You can effectively disable the asynchronous reporting feature by setting the
interface to quiet mode. All messages output from the analyzer (including those output in
response to a command line request) have the format:
X DDD:HH:MM IIII MESSAGE
X is a character indicating the message type, as shown in the table below.
DDD:HH:MM is a time-stamp indicating the day-of-year (DDD) as a number from 1 to 366, the
hour of the day (HH) as a number from 00 to 23, and the minute (MM) as a number from 00 to
59.
IIII is the 4-digit machine ID number.
MESSAGE contains warning messages, test measurements, DAS reports, variable values, etc.
The uniform nature of the output messages makes it easy for a host computer to parse them.
Analyzer Operator Manual, 02164, Rev. G
2
Definition
Print current analyzer mode
Backspace
Erase line
Recall last command
Execute last command
Execute command
Switch to computer mode
Definition
Execute command
Switch to terminal mode
Definition
Establish connection to analyzer
Disconnect from analyzer
6-15