Epa Protocol Calibration; Calibration Of Equipment; Table 7-7: Izs Or Z/S - Teledyne 6400A Instruction Manual

Sulfur dioxide analyzer
Table of Contents

Advertisement

TAI Model 6400A SO
Table 7-7: IZS or Z/S Valves Modes with Remote Contact Closure
Ext Zero CC
Contact Open
Contact Open
Contact Closed
* Calibrate only if Dynamic Calibration is enabled (see Table 9-5).

7.8 EPA Protocol Calibration

If the 6400A is to be used for EPA SLAMS monitoring, it must be calibrated in accordance with
the instructions in this section.
In order to insure that high quality, accurate measurements are obtained at all times, the 6400A
must be calibrated prior to use. A quality assurance program centered on this aspect and
including attention to the built-in warning features of the 6400A, periodic inspection, regular
zero/span checks and routine maintenance is paramount to achieving this.
In order to have a better understanding of the factors involved in assuring continuous and
reliable information from the 6400A, it is strongly recommended that Publication No. PB 273-
518 Quality Assurance Handbook for Air Pollution Measurement Systems (abbreviated, Q.A.
Handbook) be purchased from the NTIS (phone 703-605-6000). Special attention should be paid
to Section 2.9 which deals with fluorescence based SO
section is based. Specific regulations regarding the use and operation of ambient sulfur dioxide
analyzers can be found in 40 CFR 50 and 40 CFR 58. Both publications are available from the
U.S. Government Printing Office (phone 202-512-0327).

7.8.1 Calibration of Equipment

In general, calibration is the process of adjusting the gain and offset of the 6400A against some
recognized standard. The reliability and usefulness of all data derived from any analyzer depends
primarily upon its state of calibration. In this section the term dynamic calibration is used to
express a multipoint check against known standards and involves introducing gas samples of
known concentration into the instrument in order to adjust the instrument to a predetermined
sensitivity and to produce a calibration relationship. This relationship is derived from the
instrumental response to successive samples of different known concentrations. As a minimum,
three reference points and a zero point are recommended to define this relationship. The true
values of the calibration gas must be traceable to NIST-SRM's.
Analyzer Operator Manual, 02164, Rev. G
2
Ext Span CC
Operation
Contact Open
State when in SAMPLE mode, normal monitoring.
Contact Closed
Span check or calibrate
Contact Open
Zero check or calibrate
7-9
*
*