TAI Model 6400A SO
Figure 7-2: Diagram Of Calibration System
Table 7-11: EPA Expected Span Gas Concentration Procedure
Step Number
Action
1.
Press CAL-
CONC
2.
Press ENTR
3.
Press EXIT
Sample the generated concentration until the SO
Span the instrument by the following procedure:
Analyzer Operator Manual, 02164, Rev. G
2
Comment
This key sequence causes the 6400A to prompt for the
expected SO
concentration.
2
Enter the SO
span concentration value by pressing the key
2
under each digit until the expected value is set.
ENTR stores the expected SO
Returns instrument to SAMPLE mode.
response is stabilized.
2
7-15