QUICKSTART GUIDE
MODEL 3000 TA-XL-EU
Trace Oxygen Analyzer
Teledyne Analytical Instruments

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne Trace Oxygen Analyzer 3000TA-XL-EU

  • Page 1 QUICKSTART GUIDE MODEL 3000 TA-XL-EU Trace Oxygen Analyzer Teledyne Analytical Instruments...
  • Page 2 Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments...
  • Page 3: Getting Started

    Model 3000 TA-XL-EU GETTING STARTED This Quickstart Guide is designed to get you set up and operating your Teledyne Analytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss. This Quickstart Guide should be used in conjunction with the Instruction Manual that shipped with your instrument.
  • Page 4 The Standby turns off the display and outputs, but circuitry is still operating. Access Door: For access to the Micro-fuel Cell,the front panel swings open when the latch in the upper right corner of the panel is pressed all the way in. Teledyne Analytical Instruments Trace Oxygen Analyzer...
  • Page 5: Rear Panel (Equipment Interface)

    Range ID Contacts: Four separate, dedicated, range relay contacts. Low, • Medium, High, Cal. Network I/O: Serial digital communications for local network access. For future • expansion. Not implemented at this printing. Teledyne Analytical Instruments Quickstart Guide...
  • Page 6: Setup And Install

    Note: Exhaust connections must be consistent with the hazard level of the constituent gases. Check Local, State, and Federal laws, and ensure that the exhaust stream vents to an appropriately controlled area, if required. port. SAMPLE IN port. EXHAUST OUT Teledyne Analytical Instruments Trace Oxygen Analyzer...
  • Page 7: Electrical Connections

    To change between American and European fuses, remove the single retaining screw, flip fuse block over 180 degrees, and replace screw. Replace fuse as shown below. Reassemble Housing Teledyne Analytical Instruments Quickstart Guide...
  • Page 8 Install the Micro-fuel Cell The Model 3000TA-XL-EU can be equipped with either the standard B-2CXL Micro-fuel Cell or as an option, Teledyne’s Insta Trace cell. B-2CXL Micro-fuel Cell To install this cell: 1. Remove power to the instrument.
  • Page 9 10. Purge the system using sample or zero gas. 11. Power the system back up. The Insta Trace sensor will typically reach a stable low value within 30 minutes or less after installation. Teledyne Analytical Instruments Quickstart Guide...
  • Page 10: Basic Operation

    The Enter button is used to accept any new entries on the VFD screen. The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button. Teledyne Analytical Instruments Trace Oxygen Analyzer...
  • Page 11: Available Functions

    Model 3000 TA-XL-EU Available Functions The six function buttons are used to control the analyzer. Some functions have submenus that allow additional choices as shown below. Teledyne Analytical Instruments Quickstart Guide...
  • Page 12: Password Protection

    • password. If a password has been previously installed, enter the password using the TRAK/HLD Auto—Cal PSWD Logout More T E T A I Enter PWD A A A A A Enter PWD Teledyne Analytical Instruments Trace Oxygen Analyzer PSWD...
  • Page 13: Calibration Of The Analyzer

    AUTO appears blinking. Press Enter to begin zeroing. • Change Password? <ENT>=Yes <ESC>=No AUTO Zero: Settling: AUTO <ENT> To Begin #### PPM Zero Slope=#### ppm/s #### PPM Zero 4 Left=### ppm/s Teledyne Analytical Instruments Quickstart Guide zero settling. Stop when...
  • Page 14: Span Calibration

    #### ppm Zero Slope=#### ppm/s AUTO Span: Settling: AUTO <ENT> For Next Calib. Holding time Cal hold: 5 min Span Val: 000008.00 <ENT>Span <UP>Mod # Teledyne Analytical Instruments Trace Oxygen Analyzer zero settling. Stop when span settling. Stop when AUTO...
  • Page 15: Setting Alarm Parameters

    Span: Settling:MAN <ENT> For Next AUTO Calib. Holding time Cal hold: 5 min Span Val: 000008.00 <ENT>Span <UP>Mod # Mod # #### % Span Slope=#### ppm/s Teledyne Analytical Instruments Quickstart Guide span settling. Stop when ) of span value.
  • Page 16 AL—1 1000 ppm HI Dft—N Fs—N Ltch—N ppm (oxygen); value can be set from 0.00 AL–1 #### (Yes/No) (Yes/No) (Yes/No). AL–2 and then press either ▲ two times or ▼ two Dft– L—1 M—10 H—100 Mode—AUTO Teledyne Analytical Instruments Trace Oxygen Analyzer...
  • Page 17: The Analyze Function

    Analyze function. Alternatively, you can press the Analyze button at any time to return to analyzing your sample. AUTO L—1 M—10 H—100 Mode—FX/LO L—1 M—10 H—100 Mode—FX/MED L—1 M—10 H—100 Mode—FX/HI Teledyne Analytical Instruments Quickstart Guide AUTO , or to set the FX/LO FX/MED FX/HI...