TAI Model 6400A SO
Step 3 - Review the quality of the calibration:
Calibration Quality Check Procedure
Step Number Action
1.
Scroll the TEST
function menu until
the SO
displayed.
2.
Scroll the TEST
function menu until
the SO
displayed.
Step 4 - The 6400A is now ready to measure sample gas.
Analyzer Operator Manual, 02164, Rev. G
2
Comment
Typical SLOPE value is 1.0 +/- 0.3. If the value is not in this
range, check Section 7.10 or 9. If the SLOPE value is in the
SLOPE is
acceptable range the instrument will perform optimally.
2
The 6400A will display the OFFSET parameter for the SO
equation. A value between 50mV to 250mV indicates
OFFSET is
calibration in the optimal range. If the OFFSET value is
2
outside this range, check Section 7.10 and 9.