Required Calibration Test Equipment; Start Calibration - Fluke 5080A Service Manual

Hide thumbs Also See for 5080A:
Table of Contents

Advertisement

5080A
Service Manual

Required Calibration Test Equipment

Table 29 shows the test equipment necessary.
Quantity

Start Calibration

To start a calibration:
1. Push .
2. Push the CAL softkey twice.
3. Push the 5080A CAL softkey.
The CALIBRATION SWITCH on the Product rear panel can be in
either position when you begin calibration. It must be set to ENABLE
to store the correction factors into nonvolatile memory.
After you push the 5080A CAL softkey, the procedure works as follows:
1. The Product automatically programs the outputs and prompts you to make
2. The Product then goes into Operate mode, or prompts you to place it into
3. You are then prompted to type in the value read on the measurement
To redo a step, push the OPTIONS, and BACK UP STEP softkey, or
skip over a step by pressing the SKIP STEP softkey
44
Table 29. Test Equipment Required for Calibration
Manufacturer
1
Fluke
1
Fluke
Fluke
1
1
Fluke
1
Fluke
1
Fluke
1
Fluke
1
Guildline
1
Fluke
1
Fluke
1
Fluke
1
Fluke
1
Guildline
external connections to applicable measurement instruments.
Operate mode.
instrument.
Model
8508A
Reference Multimeter
5790A or 5790B
AC Measurement Standard
A40B
10 mA, 50 mA, 20 A, 2 A, 1 A, and
200 mA current shunts
742A-1k
Resistance Standard, 1 kΩ
Resistance Standard, 100 Ω
742A-100
Resistance Standard, 10 Ω
742A-10
Resistance Standard, 1 Ω
742A-1
0.01 Ω shunt
9230
PN 900394
Type N Dual Banana Adapter
792-7004
A40 Current Shunt, Adapter
742A-1M
Resistance Standard, 1 MΩ
742A-10M
Resistance Standard, 10 MΩ
9334/100M
Resistance Standard, 100 MΩ
Note
Note
Equipment

Advertisement

Table of Contents
loading

Table of Contents