Percent Oxygen Analyzer
3.3.2.4 P
ROFIBUS
There is a port for Profibus communication. The address is set in the in
the System menu through the front panel touchscreen. The master
Profibus PLC needs a GSD file to connect to the 3000RS. Request the
GSD file from customer service at TAI.
Use only commercially available Profibus connectors. Do not build your
own. It is recommended to use a straight out Profibus connector such as
Siemens
6GK15000FC10 due to size and accessibility restrictions.

3.4 Installing the Micro-Fuel Cell

The Micro-Fuel Cell is not installed in the cell block when the
instrument is shipped. Install it before the analyzer is placed in service.
Refer to the procedure described in Section 5.2 for more information.
Once it is expended, or if the cell is exposed to air for too long, the
Micro-Fuel Cell will need to be replaced. The cell could also require
replacement if the instrument has been idle for too long.
When the Micro-Fuel Cell needs to be installed or replaced, follow
the procedures in Chapter 5, Maintenance, for removing and installing
cells.

3.5 Testing the System

Before plugging the instrument into the power source:
Check the integrity and accuracy of the gas connections.
Make sure there are no leaks.
Check the integrity and accuracy of the electrical
connections. Make sure there are no exposed conductors.
Verify that the restriction device has been properly
installed (see section 3.3.1).
Check that inlet sample pressure is within the accepted
range (see section 3.3.1).
Power up the system and test it by repeating the Self-
Diagnostic Test as described in Chapter 4, Section 4.3.5.
Teledyne Analytical Instruments
P
ORT
Installation
31