Fixed Range Analysis - Teledyne TEXIO PTE6-200 Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

Operation
3. Clicking on home button
Main screen without saving the edited values.
4. Clicking on the back button
to the System screen. If any changes have been entered, a
message box will appear requesting: "Do you want to Save
the changes?" Use the Yes or No buttons to reply.
• "Yes" will save user entered values and move to the
System screen.
• "No" will discard the changes and move to System
screen.

4.6.2 Fixed Range Analysis

The auto ranging mode of the instrument can be overridden, forcing
the analyzer DC outputs to stay in a single predetermined range.
To switch from auto ranging to fixed range analysis, enter the range
function by pressing the Range button and toggle the Range Type from
AUTO to FIXED.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value) as
established by the operator for that range, the output saturates
at 1 VDC (or 20 mA). However, the digital readout and the RS-
232 output continue to read the true value of the oxygen
concentration regardless of the analog output range.
Teledyne Analytical Instruments
3000RS-P
will move the display to the
will move the display back
80