Teledyne TEXIO PTE6-200 Operating Instructions Manual page 114

Percent oxygen analyzer
Table of Contents

Advertisement

Percent Oxygen Analyzer
The current oxygen concentration (in percent) will be updated in
the Oxygen % field as will the slope as calibration proceeds.
The flowmeter display is used to show the current flow rate and the
trend in flow (fluctuating, stable, etc.). It shows the current flow values
based on scale factor. Low scale limit = 0 and the High scale limit =
100.
The Calibration status bar displays calibration progress messages as
well as any errors that may occur.
1. Message 1: "Calibration Progress" - This message displays
in the Status bar as the calibration begins.
2. Message 2: A calibration countdown displays as:
"Calibration 5 Left", "Calibration 4 Left" etc. and is based
on the currently calculated gain offset.
3. Message 3: Error messages
• "Error: Span Failed!" - Zero calibration has
• "Error: Input Still Too High" - Calibration
• "Error: Can't Span Flow Failed" - This error
5. Message 4: "Calibration done" - This indicates that the zero
calibration has successfully completed without any errors.
The Span Calibration Run screen has a Back arrow button
when pressed DURING A CALIBRATION will bring up a display
which will allow the abort the calibration without saving. Use the Yes
or No buttons to reply.
Teledyne Analytical Instruments
failed.
aborted while doing gain switching.
appears if there has been a flow failure during the
calibration.