Percent Oxygen Analyzer
.
1. The Serial Number can range from 100000 to 999999.
2. Click on the "DEL" button to delete the one character at a
time from the Edit box.
3. Click on "ENT" Button to store the values and move back to
the parent (More) screen.
4. Click on the "back arrow" button on this screen to move back
to the More screen without saving the edited value.
4.5.10.2 N
EGATIVE
The instrument can be reconfigured to show negative readings
which would be useful if the sensor output has drifted below zero. This
situation may arise after the instrument has been zeroed, as time
progresses the sensor may drift below the zero calibration setpoint.
To show negative oxygen readings on the display, toggle the edit
box from NO to YES.
1. Negative number- ON: The system will show negative values
of slope or calculation result.
2. Negative number- OFF: The system will show 0.00 when
negative values are obtained from a calculation result.
This preference is stored in non-volatile memory, so this
configuration is remembered after a power shutdown. If the instrument
is cold started, it will go back to default (not showing negative oxygen
readings).