Profibus - Teledyne 2000RS Operating Instructions Manual

Thermal conductivity analyzer
Table of Contents

Advertisement

Installation
The RS-232 protocol allows some flexibility in its implementation.
Table 3-6 lists certain RS-232 values that are required by the 2000RS
implementation.
Table 3-6: Required RS-232 Options
Parameter
Baud
9600
Byte
8 bits
Parity
None
Stop Bits
1
Message
2 seconds. When CRC opens again, zero and span are
Interval
done
3.7.4
P
ROFIBUS
There is a port for Profibus communication. The address is set in the in
the System menu through the front panel touchscreen. The master
Profibus PLC needs a GSD file to connect to the 2000RS. Request the
GSD file from customer service at TAI.
Use only commercially available Profibus connectors. Do not build your
own. It is recommended to use a straight out Profibus connector such as
Siemens
6GK15000FC10 due to size and accessibility restrictions.
Teledyne Analytical Instruments
Setting
2000RS
26