Percent Oxygen Analyzer
The following items are contained within the Zero calibration
manual mode run screen:
1. Calibration manual ENTER button
2. Oxygen % values
3. Slope values
4. Calibration status bar
5. Flow meter with flowrate and status
Except for the manual ENTER button all remaining options are the
same as for the AUTO mode. See section 4.4.1.1 for descriptions.
The Manual Calibration ENTER button is used to save the
calibration offset values and will only appear in the Manual calibration
mode.
Generally, you have a good zero when Slope is less than 0.05 %/s
for about 30 seconds. When Slope is close enough to zero, press
ENTER. In a few seconds, the screen will update. Once span settling
completes, the information is stored in the microprocessor.
The Zero Calibration Run screen has a Back arrow
pressed DURING A CALIBRATION will bring up a display which will
allow the abort the calibration without saving. Use the Yes or No
buttons to reply.