Zero Cal - Teledyne TEXIO PTE6-200 Operating Instructions Manual

Percent oxygen analyzer
Table of Contents

Advertisement

Operation
If you are using password protection, you will need to enter your
password to gain access to either of these functions. Follow the
instructions in Sections 4.5.1 to enter your password. Once you have
gained clearance to proceed, you can enter the Zero or Span function.

4.8.1 Zero Cal

The Zero button on the Menu screen is used to enter the Zero
calibration function. Zero calibration can be performed in either the
automatic or manual mode. In the automatic mode, an internal
algorithm compares consecutive readings from the sensor to determine
when the output is within the acceptable range for zero. In the manual
mode, the operator determines when the reading is within the acceptable
range for zero. Make sure the zero gas is connected to the instrument.
Make sure the Zero calibration gas source is connected to the
instrument (see Section 3.3.1). Press the Zero button on the Menu screen
to enter the Zero Calibration function.
Menu screen:
Teledyne Analytical Instruments
3000RS-P
88