Percent Oxygen Analyzer
1. Click the "DEL" button to delete the one character at a time
from the Edit box.
2. Click the "ENT" Button to store the values and move back to
the parent Span Cal screen.
3. Click the "back arrow" to move to the Span Cal screen
without saving any edited values.
Calibration Concentration Value:
To enter or check the Span Concentration of the calibration gas
touch the corresponding edit box to bring up the numerical pad.
Enter the oxygen concentration using the same procedure as for the
Cal holding time described above. When you have finished typing in the
concentration of the span gas you are using in % (20.90 if you are using
air), press ENT to save the value or the back arrow to discard the entry
and dismiss the numerical pad.
Start the Calibration:
To begin the Span calibration, touch the corresponding box on the
Span Cal screen.