Rs-232 Port; Figure 3-5: Fet Series Resistance; Table 3-5: Commands Via Rs-232 Input - Teledyne 2120XL Operating Instructions Manual

Trace nitrogen in argon analyzer
Table of Contents

Advertisement

Installation

Figure 3-5: FET Series Resistance

3.5.3 RS-232 Port

The digital signal output is a standard, full duplex RS-232 serial
communications port used to connect the analyzer to a computer,
terminal, or other digital device. It requires a standard 9-pin D
connector.
The output data is status information, in digital form, updated every
two seconds. Status is reported in the following order:
The concentration in ppm or percent
The range in use (HI, MED, LO)
The span of the range (0-100 ppm, etc)
Which alarms—if any—are disabled (AL–x DISABLED)
Which alarms—if any—are tripped (AL–x ON).
Each status output is followed by a carriage return and line feed.
Three input functions using RS-232 have been implemented to
date. They are described in Table 3-5.

Table 3-5: Commands via RS-232 Input

Command
as<enter>
Teledyne Analytical Instruments
Model 2120XL
Description
Immediately starts an autospan.