Percent Oxygen Analyzer
3. Message 3: Error messages
• "Error: Zero Failed!" - Zero calibration has
• "Error: Input Still Too High" - Calibration
• "Error: Can't Zero Flow Failed" - This error
4. Message 4: "Calibration done" - This indicates that the zero
calibration has successfully completed without any errors.
The Zero Calibration Run screen has a Back arrow button
when pressed DURING A CALIBRATION will bring up a display
which will allow the abort the calibration without saving. Use the Yes
or No buttons to reply.
• "Yes" will ABORT the calibration and move back to
the Menu screen.
• "No" will continue the calibration.
Teledyne Analytical Instruments
failed.
aborted while doing gain switching.
appears if there has been a flow failure during the
calibration.
Operation
that
91