Clicking on "Yes" will save the entered values and move to the
System screen. Clicking on "No" will discard the changes and move to
System screen.

4.5.4 Auto Cal Screen

When proper automatic valving is connected (see Chapter 3,
Installation), the analyzer can cycle itself through a sequence of steps
that automatically zero and span the instrument. The Auto-Cal screens
that follow can be used to define an automatic calibration sequence
and/or start an Auto-Cal event.
The Auto Cal function is available from the first System screen.
To enter the Auto Cal function press System and then Auto Cal. A
series of screens will appear beginning with the following screen:
Auto Cal screen has the following options,
A. Auto Cal Zero ON or OFF
B. Auto Cal Span ON or OFF
The Zero/Span schedule can be set to xxx days and xx hours before
the next auto calibration cycle will automatically occur. The Zero and
Span rows have Days and Hours Edit boxes. Clicking on the edit box
will bring up individual number pads for entering or editing the Days
and Hours setting.
Teledyne Analytical Instruments
3000RS-P