Internal Trigger; Trigger Output; 1Trigger Output Port Selection; 2Frame Sync Output On Sync Connector - Teledyne ADQ8-4X Manual

Table of Contents

Advertisement


 
#
DESCRIPTION
a
The level trigger has a hysteresis function to avoid false triggering on
noise.
b
When the signal passes below the RESET_LEVEL_CODE, the ADQ may
trigger again.
c
Trigger position is the first sample above the trigger
d
Set the trigger LEVEL_CODE.
e
Set the trigger RESET_LEVEL_CODE.
Figure 17: Trigger reset level.

4.11 Internal trigger

The internal trigger generates a periodic trigger signal. The internal trigger period is specified in number
of samples.
Note that if the internal trigger signal used as a trigger output, the actual output signal is updated at the
Data Clock rate. This may appear as jitter on the output. It is recommended to use a period that is a
multiple of the Data Clock is the internal trigger shall drive external equipment.

4.12 Trigger output

The trigger output signal is intended for triggering external equipment connected to the ADQ to build a
synchronized system. It can also be used for indicating that a trigger event occurred in the digitizer. The
trigger output is updated at the rate of the Data Clock.
The trigger output function consists of two parts:
The first part selects the source of the trigger output signal.
The second part selects the physical output port for the trigger output signal.
Note that the trigger output is the same physical TRIG connector as the external trigger input.
4.12.1 Trigger output port selection
The trigger output port is selected to one of these ports:
Trigger connector on the front panel. Note that the trigger output is the same physical TRIG con-
nector as the external trigger input.
PXIe backplane triggers,
4.12.2 Frame sync output on SYNC connector
The frame sync feature enables grouping of trigger signals into frames or blocks or lines. The name for
this feature relate to the actual application. This function can, for example, be used in scanning three-
dimensional measurements where a record is the first dimension, the trigger is the second and the
frame sync is the third dimension.
19-2302 A
2020-09-16

)*
) *
)+*
LEVEL_CODE
Section
4.9.1.
ADQ8-4X Manual
19-2302 A 2020-09-16
USER COMMAND
REF
4.10.3
SetupLevelTrigger
4.10.1
SetupLevelTrigger
4.10.3
26(46)
) *
) *