Percent Oxygen Analyzer
Zero Cal screen 1:
A. The Zero Cal screen is used to select the Zero calibration
mode - AUTO or MANUAL
B. Click the "Zero Setting" edit box to toggle between "AUTO"
and "MANUAL".
C. Click on "Touch to Start the Calibration" to start the
calibration process. Details about manual and auto calibration
run screens are detailed below.
4.8.1.1 A
M
UTO
ODE
When the Zero calibration mode is set to Auto and the "Touch to
Start the Calibration" button is pressed, the Zero Auto Run screen is
displayed.
Teledyne Analytical Instruments
1. AUTO: Auto mode will calculate the
calibration offset automatically.
2. MANUAL: Selecting manual mode, the user
must press "ENTER" to save calculated
calibration offset.
Z
EROING
Operation
89