Teledyne TEXIO PTE6-200 Operating Instructions Manual page 135

Percent oxygen analyzer
Table of Contents

Advertisement

Appendix
OPERATION WITHOUT A RESTRICTORDEVICE:
Operation without a restrictor device is not recommended as
mentioned above. A 3000T (non-touch screen version) without any flow
restrictor device was tested on 11-19-97. This results in a flow rate of
2.4 slpm at 1 psig. This is a cv of 0.023 for the standard sample system.
Same results and recommendations should apply to 3000RS-P series.
REFERENCE: FLOW_1 .XLS & FLOW _2.XLS for
information on flow rates at various pressures.
TAI PART NUMBERS
1/8" Union (ss)
1/8" LP. Restrictor R3779
1/8" Std. Restrictor R3778
CONVERSIONS:
1 PSI
1 SCFH
Teledyne Analytical Instruments
U211
Low pressure /vac. service
High pressure 5-50 psig
=
2.04 INCHES OF MERCURY (in. Hg.)
=
0.476 SLPM
3000RS-P
120

Advertisement

Table of Contents
loading

Table of Contents