Executing Methods

Not all HART® devices have the same methods available.
The function, purpose and execution of each method can
be different. Methods may include the following:
• Self-test
• Loop test
• Sensor trim
• D/A trim
Execute a method as follows:
1. Tap the method name. A screen opens with information
about the selected method. Four buttons are displayed
on the screen
• Help - Displays a description of the method
• Abort - Exits from the procedure
• OK - Accepts inputs and proceeds to the next step
• Switch App - returns to Druck DPI 620 Genii screen
(changes channel function settings without
interrupting the method procedure)
2. Some methods may make the HART® device output a
certain current
Note: A warning will appear on the screen before the devices
enters the mode.