4. Check the calibration is correct:
• Select the applicable Current (source) function(Ref:
Section 2.4 and Section 2.9)
• Apply the values that follow:
• Check the error is within limits (Ref: Table 8-3)
Source mA
06 June 2013
CH1/CH2:
Table 8-3 Current (source) error limits
Calibrator
uncertainty
±0.2
±6
±12
±18
±24
0.2, 6, 12, 18, 24
(mA)
0.00008
0.00023
0.00044
0.0065
0.0012
Druck DPI620 Genii
Permitted
DPI 620
error (mA)
0.0010
0.0016
0.0022
0.0028
0.0034