GE Measurement & Control Druck DPI 620 Genii User Manual page 53

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
Example A
Measure frequency on
CH1
Range: 0 to 50 kHz
Trigger level: 2.5V
Note (If in Auto Trigger,
trigger value shows
current auto trigger level)
1. Set the applicable
software options
2. Complete the electrical
connections
3. If necessary, change the
Trigger Setup
SETTINGS >>
AUTO TRIGGER
4. Set the values that
follow:
AUTO TRIGGER
(Enabled / Disabled)
MANUAL LEVEL
Trigger Level for
Manual Trigger only
06 June 2013
Example B
Source frequency on CH1
Range: 0 to 50 kHz
Waveform: Square
Amplitude:5.0 V
1. Set the applicable software
options
2. Complete the electrical
connections
3. If necessary, change the
Waveform Setup:
SETTINGS
4. Set the values that follow:
WAVEFORM
Square
Triangle
Sine
AMPLITUDE
Peak to Peak Amplitude
OFFSET
Only Applies to Sine and
Triangle
Druck DPI620 Genii
Page 53 of 179