Failed To Find Device - GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

User Manual
6.14

Failed to find Device

Fault
Possible
Cause
Failed
Power
to find
Supply
device
Device under
test
Circuit
Druck DPI 620
Genii
Preferences
06 June 2013
Action
Check device is turned on
Check applicable fuses
Check the supply voltage is within limits
Check the loop current is in the range
3.5mA to 24mA
Check device is HART® compatible
Connect multiple devices one at a time
Check circuit connections
Check circuit Continuity
Check correct polarity of supply to the
transmitter
Check the HART® resistor is in the correct
place in the loop
Check the HART® resistor has the correct
value
Check the Druck DPI 620 Genii is
connected to the correct points in the
loop
No external HART® resistor present, check
the internal resistor setting
External HART® resistor present, check the
DPI 620 resistor setting is OFF
Check that the CH 2 function is set to
'NONE' if the Druck DPI 620 Genii is
used as a secondary master (in parallel
with an external supply
Select the 'search 0-63' option to scan all
possible Poll Addresses to get the Poll
Address and Tag details of the
connected device
Druck DPI620 Genii
Page 101 of 179